Product Description
SYSTEM DESCRIPTION
Fully Automatic, random access, continuous loading, benchtop open analyzer for Clinical Chemistry and Immunoturbidimetric assays
ASSAY TYPES
- End point, Kinetic, Differential, Fixed time, Bichromatic and Immunoturbidimetric
- up to 999 tests, all user programmable
REAGENTS ON BOARD
Up to 72 reagents can be located into the refrigerated compartment and distribuited on:
- 4 Pharma exports fail to meet FY20 target due to curbs stanobol get steroids to gain muscle, anabolic steroids for sale reviews – edisn removable racks
- 50/20/5 ml containers can be loaded
- 8 extra positions for 5 ml containers
SAMPLE LOADING
- 4 universal, separate racks for continuos loading of samples, calibrators and controls
- each rack contains up to 15 cups and/or tubes 10-16 mm diam., 40-100 mm height
- 8 extra position for STAT samples, calibrators and controls
- STAT sample execution is allowed at any time
THROUGHPUT
Up to 420 tests/hour including I.S.E.
(280 tests/hour photometric)
READING SYSTEM
- direct reading, dual channel photometer
- 200 μl minimum reading volume
- 9 narrow band, automatically selected interferential filters 340, 380, 405, 492, 510, 546, 577, 620, 690
- halogen lamp 6V/10W
- linear from 0.0001 to 4.200 Abs.
- high resolution (0.0001 Abs.)
SAMPLING PROCESS
One mechanical arm performs all sampling operations with:
- liquid level sensing (capacitive)
- reagent pre-warming at 37° C
- automatic probe washing
- sample volume range: 2.0 μl ÷ 99 μl (0.25 μl incr.)
- reagent 1 volume range: 2.0 μl ÷ 348 μl (1 μl incr.)
- reagent 2 volume range: 2.0 μl ÷ 330 μl (1 μl incr.)
- reagent 3 volume range: 2.0 μl ÷ 330 μl (1 μl incr.)
- sample pre-dilution
- sample post-dilution
- sample post-concentration
REACTION PLATE
Holds 80 cuvettes divide into 4 racks with 20 cuvettes each and performs the following operations:
- automatic cuvette washing
- continuous monitoring of cuvette quality
- incubation temperature control at 37° C ± 0.3
COMPUTER (OPTIONAL) MINIMUM REQUIREMENTS
- CPU: Intel or AMD
- MEMORY: Ram 2 Gbyte
- HARD DISK: 250 Gbyte
- MONITOR: 17“
- COMMUNICATION: USB and/or RS232 port
- HOST COMMUNICATION: RS232 or Ethernet
- OPERATIVE SYSTEM: WINDOWS XP – VISTA – 7 (32 bit only)
OPTIONS
- Direct potentiometry I.S.E. module for Na+, K+, CI- (140 tests/hour)
- Positive barcode readers for reagents and samples (code 128, codebar, code 2 of 5 interleaved, code 39)
INSTALLATION REQUIREMENTS
- Power Supply: 90 ÷ 250 VAC, 47 ÷ 63 Hz
- Electric consumption: 386 Watt
- Water Consumption: 1.5 l/h
DIMENSIONS, WEIGHT & ENVIRONMENT
- Analytical Module:
- Height: 50 cm
- Width: 68 cm
- Lenght: 105 cm
- Weight: 49 Kg
- Room Temperature: 18° C ÷ 30° C
- Relative Humidity: 20% ÷ 85%